Introduction:

Thermo Scientific Orbitrap Exploris Mass Spectrometer is a high resolution mass spectrometer that is capable of collecting high quality MS and tandem MS data for both non-targeted and targeted analysis. The fragmentation techniques of HCD facilitate MS/MS fragmentation which can apply in environmental analysis and metabolomic research.

 

Specifications:

Thermo Scientific TRACE 1600 Series GC

  • Carrier gas: Helium
  • Reagent gas: Methane (for CI)
  • Split/Splitless injector (SSL)
  • Oven Operation Temperature range: ambient+3°C to 450°C   
  • Oven Temperature set point resolution: 0.1°C 
  • Oven cool down: 450°C to 50°C in less than 4 minutes
  • Split ratio: up to 12500:1
  • Split flow: OFF or from 0-1250 mL/min in 0.1 mL/min increments
  • Pressure range: 0-150 psi
  • Injector Maximum temperature: 400°C

 

TriPlus RSH SMART Advanced Autosampler

Perform the change of the injection tool automatically for unattended sequence with different samplling techniques, and supporting the automation of multi-setep sample preparation workflows.

  • Incubator/agitator
    • Capacity 6 x 20 mL vials (compatible with 10 mL vials with adapters)
    • 30-200°C temperature range
    • 250-750 rpm agitation speed

 

  • Headspace Sampling
    • Compatible with 2, 10 and 20 mL vials
    • Syringe temperature: OFF or 40°C to 150°C in 1°C steps
    • Incubation oven: 30-200°C temperature range, in 1°C steps; 250-750 rpm agitation speed
    • Incubation time: 0.1 to 600.0 min in 0.1 min increments

     

  • Solid-phase microextraction (SPME)
    • Incubation oven: 30-200°C temperature range, in 1°C steps; 250-750 rpm agitation speed
    • Incubation time: 0.1 to 600.0 min in 0.1 min increments

Thermo Fisher Orbitrap Exploris Mass Spectrometer

  • Ion Source:
    • Electron ionization (EI)
    • Chemical ionization (CI): Postiive Ion Chemical ionization (PCI) and Negative Ion Chemical Ionization (NCI)
  • Scan modes: MSn (n = 1-10)
  • Fragmentation techniques in MS/MS:
    • High-energy collision dissociation (HCD)
  • Mass analyzer:
    • Orbitrap MS
  • Mass range:
    • m/z 30-3,000
  • Resolution: 60,000 FWHM at m/z 200
  • Scan rate: 40 Hz at resolution setting 7,500 at m/z 200
  • Mass accuracy:
    • Less than 3ppm RMS using external calibration
    • Less than 1ppm RMS using internal calibration

Applications:

  • Environmental and Food Analysis
  • Metabolomics

Software:

  • Compound Discoverer (For Non-Target Analysis)
  • TraceFinder (For Quantitative Analysis)
  • Progenesis QI (Waters)
 

Note to users:

First-time users must be trained by equipment-in-charge before they can use the instrument independently.
Samples should be desalted and purified prior to MS analysis to enhance the detection sensitivity and minimize contamination.

Booking:

Please visit our University Research Facility Management System (URFMS) website for registration, training request and booking arrangement.

 

Contact Information:

Location: W711, Block W, The Hong Kong Polytechnic University
Equipment-in-charge: Dr. Chi-Hang CHOW/ Sirius TSE (Office: TU503)
Email: chihang.chow@polyu.edu.hk / sirius.tse@polyu.edu.hk
Telephone: 3400 8813/ 3400 8878

Your browser is not the latest version. If you continue to browse our website, Some pages may not function properly.

You are recommended to upgrade to a newer version or switch to a different browser. A list of the web browsers that we support can be found here